SIGMA II-3

Have questions about our products? Check out our FAQ section.

The SIGMA II-3 Multiplace Hyperbaric System is designed to accommodate three seated patients simultaneously. The SIGMA II-3 delivers exceptional performance for hospitals with moderate patient treatment requirements. The SIGMA II-3 bridges the gap between the clinical limitations of the monoplace system and the major investment required for larger multiplace chambers including operating cost savings and space requirements.

Perry Baromedical, hyperbaric oxygen therapy, hbot, hyperbaric chambers, hyperbaric chamber, wound healing, oxygen therapy

Flexibility

The SIGMA II-3 features a spacious 40-inch internal diameter acrylic treatment compartment that allows for increased patient accessibility and monitoring. The ability to treat three patients seated comfortably is an important SIGMA II-3 distinction. Pressurized with air instead of oxygen, the SIGMA II-3 utilizes a built-in breathing system (BIBS) that gives the operator the ability to independently adjust each patient’s oxygen supply.

Patient Comfort

The SIGMA II-3’s spacious interior increases patient treatment acceptance by reducing the likelihood of confinement anxiety. Patient comfort is also enhanced by the cushioned, reclining chairs. The acrylic compartment allows the patient excellent vision and the ability to enjoy entertainment (TV/Video system) during the treatment period.

Safety

The SIGMA II-3 is a Class A hyperbaric chamber as defined by the National Fire Protection Association (NFPA). The standard fire suppression system is activated at the control console. Patients are supplied with breathing oxygen gases separate from the system environment. A standard oxygen monitor checks the concentration of oxygen within the compartment.

Patient Care

The SIGMA II-3 also allows for medical monitoring of critically ill patients. The optional medical lock is a convenient way to pass refreshments or medical supplies in and out of the system.

Cost-Effectiveness

The SIGMA II-3 reduces staffing requirements per patient while increasing the number of patients that can be treated simultaneously. The built-in breathing system (BIBS) provides optimal oxygen consumption per patient.

Building Service Requirements

The following requirements must be met for proper operation of the SIGMA II-3 Multiplace Hyperbaric System.

  • Building must adhere to construction requirements specified in NFPA 99, Health Care Facilities.
  • Optimal patient comfort is achieved if the room is maintained at 68°-72° F (20°-22.2° C).
  • Breathing Quality Air: Main Air - minimum pressure of 60 psi at 40 scfm (4.13 bar @ 1133 lpm). Stand-by Air - minimum pressure of 50 psi at 7.4 scfm (3.4 bar @ 210 lpm).
  • Oxygen Supply: Minimum pressure of 50 psi @ 7.4 scfm (3.4 bar @ 210 lpm). Pressure gauge and shut-off valve installed in accessible location. Line and all components installed in accordance with NFPA specifications.
  • Stand-by Air Supply: A minimum of 210 lpm if breathing quality air at a minimum pressure of 50 psig must be supplied to the chamber as stand-by (emergency breathing) air. The air must be taken from a source other than the main air supply, either high-pressure cylinders or from existing hospital air supplies if a separate dedicated compressor system is used.
  • Exhaust Line: 1½” nominal diameter copper (minimum). The exhaust line should be run to the exterior of the building no less than eight feet above the chamber level in an area where no potential source of ignition or fuel exists. The exhaust line should be installed so as to obviate entry of rain or foreign objects.
  • Fire Suppression System: (A) Fire suppression deluge water - flow of 80 gpm @ minimum pressure of 110 psi (B) Hand line water, from a separate, independent source - flow of 10 gpm @ minimum pressure of 85 psig.
  • Electrical Service: 110 VAC outlet for communication system power supply and maintaining charge on 12 VDC backup battery.
  • Chamber Ground: Chamber grounding should be in accordance with NFPA 99 , NFPA 70, National Electrical Code, grounding and local electrical codes. Additional grounding instructions shall be provided as required.
  • The SIGMA II-3 is supplied with all necessary hoses and fittings to easily connect to the customer-provided services described above. All building services should terminate within three feet of the foot of the hyperbaric system.

Technical Specifications

Dimensions
Overall Length
129 inches (3.28meters)
Overall Width (w/out optional medlock and cosmetic covers removed)
46 inches (1.17 meters)
Overall Width (w/optional medlock)
58.5 inches (1.49 meters)
Overall Height
60.5 inches (1.54 meters)
Internal Diameter (acrylic)
40 inches (1.02 meters)
Internal Length (chamber)
126 inches (3.20 meters)
Internal Volume (treatment compartment)
86 cubic ft (2.43 cubic meters)
Weight
5,000 lbs (2,268 kg)
Warranty
1 year
Performance
Maximum Operating Pressure
30 psi/3 ATA (2.07 bar)
Design Temperature Range
32°-125° F (0°-52° C)
Oxygen Supply
50-70 psi @ 210 lpm (3.4 - 4.8 bar)
Primary Air Supply
60-100 psi @ 40 scfm (4.1 - 6.9 bar)
Standby Air Pressure
50-70 psi @ 210 lpm (3.4 – 4.8 bar)
Fire Suppression Deluge
80 gpm @ 110 psi (303 lpm @ 7.6 bar)
Fire Suppression Hand Held
10 gpm @ 85 psi (37.9 1pm @ 5.9 bar)
Ventilation Rate
(4.4-13.6 scfm) 125-385 lpm
Pressure Change Rate
1-5 psi/min. (.07-.35 bar/min.)
Emergency Depressurization Rate
0.25 psi/sec. (.017bar/sec.) Approximate rate